PN FBC-0059
September 2013, Rev. 1
© 2013 Fluke Corporation. All rights reserved. Specifications are subject to change without notice.
All product names are trademarks of their respective companies.
451P
Ion Chamber Survey Meter
Operators Manual
Seitenansicht 0
1 2 3 4 5 6 ... 29 30

Inhaltsverzeichnis

Seite 1 - Operators Manual

PN FBC-0059 September 2013, Rev. 1 © 2013 Fluke Corporation. All rights reserved. Specifications are subject to change without notice. All product na

Seite 2

451P Operators Manual vi

Seite 3 - Notices

1 Introduction The Model 451P Ion Chamber Survey Meter (the Survey Meter) is a hand-held, pressurized, battery operated meter designed to measure ga

Seite 4 - Manufacturing Location

451P Operators Manual 2 Safety Information A Warning identifies conditions and procedures that are dangerous to the user. A Caution identifies conditi

Seite 5 - Table of Contents

Ion Chamber Survey Meter Safety Information 3 • The battery door must be closed and locked before you operate the Product. • Replace the batteri

Seite 6

451P Operators Manual 4 • Each person who uses a product that senses and measures ionizing radiation must: o Be approved to do work where there is r

Seite 7 - List of Tables

Ion Chamber Survey Meter Features 5 Features The following sections describe the features of the Survey Meter. Overrange Rate If the instantaneous

Seite 8

451P Operators Manual 6 Specifications Radiation Detected ... Beta above 1 MeV & gamma above 25 KeV Op

Seite 9 - List of Figures

Ion Chamber Survey Meter Specifications 7 Environmental ... Operating Temperature Range: -4 °F

Seite 10

451P Operators Manual 8 Dose equivalent units were determined using conversion coefficients at H*(10), found in I CRU 47, April, 1992.00.20.40.60.811.

Seite 11 - Intended Use

Ion Chamber Survey Meter Specifications 9 1.020°40°60°80°90°100°120°140°180°137Cs.6.7.8.9.5.4.3.2.10°38 keV160° hms04.eps Figure 4. Survey Meter A

Seite 12 -  On/Off button

Warranty and Product Support Fluke Biomedical warrants this instrument against defects in materials and workmanship for one full year from the date

Seite 13 - Ion Chamber Survey Meter

451P Operators Manual 10 Receiving Inspection Upon receipt of the Survey Meter: 1. Check the shipping cartons(s) and their contents for in-shipment d

Seite 14

Ion Chamber Survey Meter Operation 11 MODE Button To configure the MODE button to the opposite function without a RS-232 connection, use the follo

Seite 15 - Features

451P Operators Manual 12 Self Test When the Survey Meter is first turned on, it runs through a functional self-test procedure. During this self-test,

Seite 16 - Specifications

Ion Chamber Survey Meter Operation 13 RS-232 hms03.eps Figure 5. Location of RS-232 port Operational check Fluke recommends an operational check u

Seite 17

451P Operators Manual 14 Theory of Operation  Caution To prevent damage to the Product, do not open or disassemble the Product. The ionization chambe

Seite 18 - Relative Response

Ion Chamber Survey Meter Theory of Operation 15 The bar graph display is a digital presentation, programmed to appear as a linear analog meter dis

Seite 19

451P Operators Manual 16 The operation portion of the firmware performs all of the control functions needed to read and control the electrometer and r

Seite 20 - Operation

Ion Chamber Survey Meter Maintenance 17 Storage Storage of Fluke Biomedical instruments must comply with Level B storage requirements as outlined

Seite 21

451P Operators Manual 18 The LOW BAT message will appear on the display approximately 6 hours prior to the Survey Meter becoming inoperable. To ensure

Seite 22 -  Warning

Ion Chamber Survey Meter Maintenance 19 Replaceable Parts Information Fluke maintains a complete inventory of all normal replaceable parts. To pla

Seite 23

Contents (continued) Notices All Rights Reserved  Copyright 2013, Fluke Biomedical. No part of this publication may be reproduced, transmitted,

Seite 24 - Theory of Operation

451P Operators Manual 20

Seite 25

To ensure the accuracy of the Product is maintained at a high level, Fluke Biomedical recommends the product be calibrated at least once every 12 mo

Seite 26 - Maintenance

i Table of Contents Title Page Introduction ... 1 Inten

Seite 27 -  Warning

451P Operators Manual ii

Seite 28 - Figure 6. Battery Access

iii List of Tables Table Title Page 1. Symbols ...

Seite 29

451P Operators Manual iv

Seite 30

v List of Figures Figure Title Page 1. Fluke 451P Ion Chamber Survey Meter .............. 4 2. Model 451P